#Product Trends

Angstrom release an upgrade for model of MRI-300

Angstrom release an upgrade for model of MRI-300

Introduction

Magnetic Resonance Imaging MRI-300 is specially designed for living animal’s observation. it provides high contrast images and accurate messages in vivo intuitively, equipped with permanent magnet. As a powerful and non-destructive MRI instrument, this system is widely used in life sciences to study tissue structure in vivo and their contrast message dispense with special maintenance cost....

Applications

Doing research related to nuclear magnetic imaging, hard ware structure and etc.

Magnetic resonance imaging experiments, 2D/3D imaging research

Imaging research with living sample (mice, plants and etc.)

Specifications

Magnet: permanent,1.45±0.05T:

H resonant frequency: 61.7±2MHz

Magnet Pole diameter: ϕ200mm

Sample field size: ϕ45mm×H80mm

Magnet field uniformity:< 8ppm

Magnet temperature control: dual system of nonlinear precise temperature control, temperature controllable from 25°C to 35°C ; temperature control accuracy ±0.005°C ; temperature display accuracy 0.001°C

Signal to noise ratio (SNR): 55dB, Degree of distortion: <1%

Graphics resolution: max. 512x512x128(3D)

Magnetic field stability: Larmor frequency variations 100Hz/Hr

Space resolution: 0.15mm, max. 0.08mm

Gradient field Intensity: 1.1 mT/m

Max gradient magnet field: 200mT/m.s

Graphics linearity: >98% in X, Y and Z direction

Absolute resolution: <0.08mm

Power supply: 110V/220V, 60/50Hz

Work temperature: 10~35°C

Environment humidity: 30~70%

Dimensions (L, W, H): Main unit 500mm×500mm×700mm.

Details

  • Stoughton, MA 02072, USA
  • Angstrom Advanced

    Keywords