#Product Trends
High-Temperature Fixed Point (1100 to 2500 ºC) Black body Furnace
•Realizing high-temperature fixed point calibration for radiation
thermometers by utilizing metal-carbon eutectic fixed points.
•Enable to fixed point calibration of standard radiation thermometer
(0.65μm) by blackbody cavity of aperture diameter of 3mm.
•Prepare 7 kinds of fixed point cells from 1100 to 2500ºC (By
exchangeable fixed point cells, single furnace has brought
calibration of 7 fixed points.)
•Enable the power-saving feature by using cylindrical sheet heating
element made of carbon-carbon fiber composite materials.
•Achieve long-term stability by employing fiber optic type radiation
thermometer (0.9μm) as temperature control sensor.
Fixed Point Metals : Model IR-R80
Copper (Cu: 1085ºC) : IR-80CU
Iron - Carbon (Fe-C: 1153ºC) : IR-80FE
Cobalt - Carbon (Co-C: 1324ºC) : IR-80CO
Palladium - Carbon (Pd-C: 1492ºC) : IR-80PD
Platinum - Carbon (Pt-C: 1738ºC) : IR-80PT
Ruthenium - Carbon (Ru-C: 1953ºC) : IR-80RU
Rhenium - Carbon (Re-C: 2474ºC) : IR-80RE