#Product Trends
Carbon Dioxide Monitor and Data Logger
The AQM-102 is designed to measure CO2 concentration,
air temperature and relative humidity with visible and audible
alarms. This CO2 monitor is an ideal instrument for indoor
air quality (IAQ) diagnosis and HVAC system performance
verification.
Carbon dioxide (CO2) is a gaseous component of the earth’s
atmosphere. The concentration of CO2 in natural ambient
air is about 0.04% or 400 ppm. With each breath, humans
convert oxygen (O2) into carbon dioxide (CO2). Although
carbon dioxide is invisible and odorless, increased CO2
levels can cause headaches, drowsiness and a loss of the
ability to concentrate or worse.
SPECIFICATIONS
CO2
Measurement Range: 0 to 9999 ppm
Accuracy: ±30 ppm, + 3% of rdg
Resolution: 1 ppm
Temperature
Range: -40 to 85ºC (-40 to 185ºF)
Accuracy: ±1.2ºC (2.2ºF)
Resolution: 0.1ºC (0.1ºF)
Relative Humidity
Range: 0.0 to 90% RH
Accuracy: ±3% RH [at 25ºC (77ºF), from 10 to 90 RH],
±5% RH beyond 10%
Memory Capacity: 32,000 readings
Storage Temperature: -40 to 70ºC (-40 to 158ºF)
Interface: USB
Dimensions: 110 H x 108 W x 28 mm D
(4.33 x 4.25 x 1.10")
Weight: 165 g (5.8 oz)
Non-Enforced Reference Levels*
250 to 350 ppm – Background (normal) outdoor air level
350 to 1000 ppm – Typical level in occupied spaces with
good air exchange
1000 to 2000 ppm – Level associated with complaints of
drowsiness and poor air
2000 to 5000 ppm – Levels associated with headaches,
sleepiness and stagnant, stale,
stuffy air; poor concentration, loss of
attention, increased heart rate and
slight nausea may also be present
>5000 ppm – Exposure may lead to serious oxygen
deprivation resulting in permanent
brain damage, coma and even death