PowerHV- Shielding Room ( PD free test site)

Shielding Room

Electromagetic field and line-conducted interference needs to be eliminated in order to measure low-intensity partial discharges with a PD background noise level (≤ 1 pC) in accordance with IEC 60270.

The effect of electromagetic field interference caused by radio signals, radar stations, welding arcs, etc. can be limited by a shielded room, also known as a Faraday cage.

PowerHV supply shielding room for Routine and type testing of cable and cable accessories.

The design and size of the shielded rooms will be determined depending on the specific application. They are separated into a test and control room. Furthermore they are equipped with a PD-free lighting system, emergency lighting, and security loop. There is the option of equipping the room with air-conditioning and a crane.