#Product Trends
Respirator Bacterial Filtration Efficiency (BFE) Tester
Equipment for testing of the bacterial filtration efficiency of surgical masks
Features:
1. Negative pressure test system to ensure the safety of operators;
2. Built - in peristaltic pump in the negative pressure cabinet, A, B, four Andersen;
3. The peristaltic pump flow can be set;
4. The flow rate of night spray of special microorganism aerosol generator can be set;
5. Good atomize effect;
6. Embedded high-speed industrial microcomputer control;
7. 10.4 inch industrial high brightness color touch screen;
8. USB interface, support U disk data storage;
9. Cabinet built-in high brightness lights;
10. Front opening and closing glass door, convenient for experimenters to observe and operate;
Specifications:
Main Specifications Parameter range Resolution Accuracy
A Sampling Flow 28.3L/min 0.1L/min ±2.5%
B Sampling Flow 28.3L/min 0.1L/min ±2.5%
Spray Flow (8~10)L/min 0.1L/min ±2.5%
Peristatic Pump Flow (0.006~3.0)ml/min 0.001ml/min ±2.5%
A Road Flow Meter (-20~0)KPa 0.01KPa ±2.5%
B Road Flow Meter (-20~0)KPa 0.01KPa ±2.5%
Pressure Before Spray Flowmeter (0~300)KPa 0.1KPa ±2.5%
Environmental Temperature (-40~99)℃ 0.1℃ ±2.5%
Negative Pressure In Aerosol Chamber (-90~-120)Pa 0.1Pa ±2.0%
Negative Pressure Of Cabinet -50~-200Pa
Data Storage Capacity >100000 Groups
Vortex Mixer Test Tube Specification and Quantity Φ16×150 mm test tubes, Eight
Characteristics of High Efficiency Air Filter The filtration efficiency of particles above 0.3 um ≥99.99%
Median Mass Diameter Of Aerosol Generator 3.0± 0.3μm,Geometric standard deviation ≤1.5
Double Route 4 Andersen Level I >7μm,Ⅱ4.7-7μm,
Level Ⅲ 3.3-4.7μm,Level Ⅳ 2.1-3.3μm,
Level Ⅴ 1.1-2.1μm,Level Ⅵ 0.6-1.1μm
Aerosol Chamber Size(L*D*T) 600*85*30mm
Total Particles of Positive Quality Control Sampler 2200±500cfu
Ventilation flow of negative pressure cabinet ≥5m3/min
Negative Cabinet Door Size(W*D) 1000×730mm
Mainframe Size(W*D*H) 1180×650×1300mm
Support Dimensions(W*D*H) 1180×650×600mm,Adjustable height within 10 cm
Instrument Noise <65dB(A)
Weight About 150kg
Power Supply AC220V±10%,50Hz
Power Dissipation <1500W