FAT finished 1 May 2018

On 1st of May 2018 FAT was finished. The equipment included a separator unit, a CIP-station 2x200 L and a 50 L freeze-dryer.

On the first day we checked the documentation for every piece of equipment, checked the equipment components presence and dimensions (for the CIP station and the freeze-dryer) and electric parts presence according to the P&ID (for the separator).

 

On the second day of FAT we did more tests:

- electrical wiring integration test for the CIP station and the freeze-dryer

- components integration test for the separator

- condencer cooling verification test for the freeze-dryer

- vacuum performance verification test for the freeze-dryer

On the third day of FAT we conducted the following tests:

- digital/analog i/o functional test and electrical safety integration test for the CIP station

- components integration test and electrical wiring integration test for the separator

- shelf temperature performance verification test and shelf heat-up rate & max temperature test for the freeze-dryer

On the 4th day of FAT we conducted the following tests:

- leakage test, control acceptance test and cleaning performance test for the CIP-station

- CIP coverage and SIP effectiveness test for the freeze-dryer

- electrical powers integration test, electrical safety integration test and spray cleaning acceptance test for the separator

On the 5th day of FAT the following tests were finished:

- ice condenser capacity and defrosting test for the freeze-dryer (it lasts 24 hours)

- digital/analog i/o functional test (separator)

From 30 April to 1 May we did leakage test, SIP test and control acceptance test for the separator and finished FAT on 1 May.